LIMS Filter
Another useful feature of the CAL3K Calorimeters
The LIMS output on the D1 port can be filtered to EXCLUDE certain types of results. the exclusion must be enabled in the SYSTEM DEFINITION (accessed via any other port).
SYSTEM DEFINITION : 'NO LIMS CALIBR.'
Setting this option excludes any calibration (and simulation) from the LIMS output.
SYSTEM DEFINITION : 'NO LIMS FAULTS'
Setting this option excludes :
Miss Fires | The sample has not ignited |
Fire Warning | The Vessel firings have exceeded the warning limit |
Inspect Vessel | The Vessel must be inspected |
Suspect | The determination is suspect |
Vessel Abort | The determination was aborted due to a vessel fault |
Manual Abort | The determination was aborted from the keyboard |
Manual Fired | The determination was manually fired |
NOTE : The filter option applies to the LIMS output only. Data outputs are NOT filtered.
This feature applies to the following CAL3K Systems : CAL3K-AP , CAL3K-A ,CAL3K-F