Sulphur Correction

The following information applies to the : CAL3K-AP , CAL3K-A , CAL3K-F.


The sulphur correction is required if the sample contains sulphur. In the presence of high pressure oxygen some of the sulphur will be converted to SO2 and some heat will be released in the conversion process. This extra heat release is removed with the sulphur correction. The sulphur correction is proportional to the sample mass and sulphur content. The sulphur correction can be set to zero.

Type: SULPHUR CORRECT, Enter, value, Enter

CAL3K Keyboard Commands - Escape Key | DDS Calorimeters Exit, Clear command entry
CAL3K Keyboard Commands - S Key | DDS CalorimetersCAL3K Keyboard Commands - U Key | DDS Calorimeters Start of command, Sulphur Correction
CAL3K Keyboard Commands - Enter Key | DDS Calorimeters Accept Command
CAL3K Keyboard Commands - 0 Key | DDS CalorimetersCAL3K Keyboard Commands - Fullstop Key | DDS CalorimetersCAL3K Keyboard Commands - 1 Key | DDS Calorimeters Enter sulphur value
CAL3K Keyboard Commands - Enter Key | DDS Calorimeters Accept Value

The Default setting is 0.0MJ/Kg

The entry of the sulphur correction is password protected (default).

The Sulphur correction is NOT used during calibration.

Sulphur Correction - Online Help | DDS Calorimeters
Sulphur Correction - Online Help | DDS Calorimeters
Sulphur Correction - Online Help | DDS Calorimeters
Sulphur Correction - Online Help | DDS Calorimeters
Sulphur Correction - Online Help | DDS Calorimeters

ZERO THE PRESSURE TRANSDUCER

The pressure transducer which monitors the vessel pressure for Filling and oxygen leakage needs to be offset adjusted so that it reads ZERO at atmospheric pressure.

Learn More

COMPENSATION

The CAL3K calorimeter range has two main compensations: Firing Wire and Cotton Compensation & Sulphur Compensation. Sulphur Compensation : Compensates for the formation of sulphuric acid. Entry in Kj/g. Default : 0Kj/g. It is disabled during calibration and calibration check.

Learn More

COTTON & WIRE CORRECTION

The sample is ignited by sending a current through the firing wire. This makes the wire glow, which in turn ignites the cotton thread, which in turn ignites the sample. The cotton and wire correction compensates for the electrical firing wire energy and the energy of the burned cotton. The default setting is 0.08°C.

Learn More